Hi Flow
Sampler Introduction
8 Instruction 0055-9017
1.6 Technical Data
TABLE 1-1. TECHNICAL SPECIFICATIONS
Specification Description
Information
Displayed
Date and Time
Battery voltage
Leak rate in cfm
Sampling flow rate in cfm
Leak concentration in ppm or % by volume
Background gas concentration in ppm or % by
volume
Percent difference between leak rate
measurements #1 and #2
Display 8 line by 20 character LCD
Pushbutton
Controls
I/O (Enter)
(Up Arrow)
(Down Arrow)
ESC (Escape)
Communication Three DB9 connectors providing serial data transfer at 115200 baud to a personal
computer, or other peripheral device
Measured Values
Sampling flow rate
Battery voltage
Sample gas concentration
Background gas concentration
Calculated Values Leak concentration corrected for background gas level
Leak rate
Percent difference between leak rate measurements #1 and #2
Measurable Leak
Rate
0.05 to 8.00 SCFM (1.42 to 226 LPM)
0.05 to 6.00 SCFM (1.42 to 170 LPM)
Accuracy
Calculated Leak Rate:
±
10% of reading
Temperature Operating: 0 to 50 °C (32 to 122 °F) Storage: 40 to 60 °C (40 to 140 °F)
Humidity 5 to 95% RH (non-condensing)
Sampling Flow
Rate
Maximum .................................... 10.5 SCFM (297 LPM) at full battery charge
Operating Flow Points ................. Initial flow 10 SCFM (283 LPM). Second flow 8
SCFM (226 LPM). (The second flow rate is 25% of the
initial flow)
Measurement Method ................ Differential pressure across restriction
Accuracy ...................................... ±5% of reading
Natural Gas Sensor
Detection Method ....................... Catalytic oxidation / Thermal conductivity
Range: Catalytic oxidation.......... 0 to 5% by volume methane
Range: Thermal conductivity ..... 5 to 100% by volume methane
Accuracy ...................................... ±5% of reading or 0.02 % methane, whichever is greater
Battery Voltage ............... 4.8 V, max.
Recharge Time ... 8 to 10 hrs
Type .......... Intrinsically Safe NiMH rechargeable pack
Run Time ... >4.5 hours continuous operation @ 20 °C
(68 °F)
Memory Stores up to 1000 individual test parameters
Dimensions 18L x 12W x 7H inches (457 x 305 x178 mm)
Weight 20 lbs (0.9 kg)
Agency Approvals Designed to be intrinsically safe for use in hazardous locations Class I, Division 1, Groups
A, B, C & D in North America. CAN/CSA-C22.2 No. 157 (June 1992). ANSI (June 27,
2002)/UL913-2002. Intrinsically Safe Apparatus and Associated Apparatus for use in
Class I, Division 1, Hazardous (Classified) Locations. CE Mark See section 8 for
Declaration of Conformity statement.