HI FLOW
®
Sampler
For Natural Gas
Leak Rate Measurement
Instruction 0055-9017
Operation and Maintenance
Rev. 6 – March 2012
Product Leadership • Training • Service • Reliability
Exclusive Distributor
Przeglądanie stron 0
1 2 3 4 5 6 ... 61 62

Podsumowanie treści

Strona 1 - Sampler

HI FLOW® Sampler For Natural Gas Leak Rate Measurement Instruction 0055-9017 Operation and Maintenance Rev. 6 – March 2012 Produ

Strona 2 - WARRANTY

Operation Hi Flow Sampler 10 Instruction 0055-9017 2.3 Sampling Hose and Attachment Connections Connect the sampling hose to the instrument’s lea

Strona 3 - Table of Contents

Hi Flow Sampler Operation Instruction 0055-9017 11 Figure 2-3. Grounding Clamp 2.5 Attachments A variety of attachments are available for the

Strona 4

Operation Hi Flow Sampler 12 Instruction 0055-9017 2.5.2 Beveled Nozzle Tool (6.5" and 24") For simple pin-hole type leaks, position t

Strona 5 - 1 Introduction

Hi Flow Sampler Operation Instruction 0055-9017 13 2.5.4 Bellows Tool This tool is useful in capturing leaks in valve stems and small fittings.

Strona 6 - Sampler?

Operation Hi Flow Sampler 14 Instruction 0055-9017 After the banner screen is displayed, the sensors are automatically zeroed to ambient conditio

Strona 7 - Figure 1-1. Control Unit

Hi Flow Sampler Operation Instruction 0055-9017 15 3. Press ESC to return to the Main Screen. Refer to Sections 2.8.1 and 2.8.2 for detailed de

Strona 8 - 1.6 Technical Data

Operation Hi Flow Sampler 16 Instruction 0055-9017 2.8.2 Expanded Main Mode Date and Time: Date displayed as mm/dd/yy, while time is di

Strona 9 - 2 Operation

Hi Flow Sampler Operation Instruction 0055-9017 17 2.9 Menu System Selecting Menu from either the Basic or Expanded Main Screen provides the use

Strona 10 - 2.4 Grounding

Operation Hi Flow Sampler 18 Instruction 0055-9017 Figure 2-10. Expanded Menu System 2.10 Battery Status Battery status is displayed as a singl

Strona 11 - 2.5 Attachments

Hi Flow Sampler Operation Instruction 0055-9017 19 2.11 Calibration Menu Selecting Calibration from the Menu System allows the user to either: 1

Strona 12 - 2.5.3 Capture Bag

WARRANTY Bacharach, Inc. warrants to Buyer that at the time of delivery this Product will be free from defects in material and manufac-ture and wi

Strona 13 - 2.6 Turning ON the Hi Flow

Operation Hi Flow Sampler 20 Instruction 0055-9017 2.14.1 Automatic 2-Stage Measurement Mode In the Automatic 2-Stage Mode the unit performs a le

Strona 14 - 2.7 Turning OFF the Hi Flow

Hi Flow Sampler Operation Instruction 0055-9017 21 2.14.2 Manual Measurement Mode 2.14.2.1 Manual 2-Stage Mode In the Manual 2-Stage mode the u

Strona 15 - 2.8.1 Basic Main Mode

Operation Hi Flow Sampler 22 Instruction 0055-9017 2. Highlight Manual 1-Stage, and then press the I/O ↵ key. 3. Press ESC to return to the Ma

Strona 16 - 2.8.2 Expanded Main Mode

Hi Flow Sampler Operation Instruction 0055-9017 23 c. Click the Connect using drop-down menu and choose the COM port to which the Hi Flow is conn

Strona 17 - 2.9 Menu System

Operation Hi Flow Sampler 24 Instruction 0055-9017 2.15.2 View Records View the records stored in memory on the unit’s LCD screen as follows: 1.

Strona 18 - 2.10 Battery Status

Hi Flow Sampler Operation Instruction 0055-9017 25 2.16 Importing Saved Data into a Spreadsheet Test records that were downloaded to a personal c

Strona 19 - 2.11 Calibration Menu

Operation Hi Flow Sampler 26 Instruction 0055-9017 TABLE 2-2. TYPICAL SPREADSHEET Row 1: Column Headings Row 2: Record 1, Typical Test Automat

Strona 20 - Operation Hi Flow

Hi Flow Sampler Operation Instruction 0055-9017 27 2.17 Access Test IDs Note: Access Test IDs can only be selected if operating in the Expande

Strona 21 - Sampler Operation

Operation Hi Flow Sampler 28 Instruction 0055-9017 2.17.2 Edit Test IDs Edit a previously entered Test ID or to create a new one as follows: 1.

Strona 22 - 2.15 Access Records

Hi Flow Sampler Operation Instruction 0055-9017 29 1. Install serial data cable P/N 104-4027 (straight through, 6 foot, DB9 male to DB9 female) be

Strona 23

Hi Flow Sampler Table of Contents Instruction 0055-9017 3 Table of Contents 1 Introduction ...

Strona 24 - 2.15.3 Erase Records

Operation Hi Flow Sampler 30 Instruction 0055-9017 Observe that the Hi Flow LCD displays the message “Sending record x of y @ 115.2kbps,” where “

Strona 25

Hi Flow Sampler Operation Instruction 0055-9017 31 2.18 Making a Leak Rate Measurement Before making a leak rate measurement, the user must decid

Strona 26 - Row 3: Record 2

Operation Hi Flow Sampler 32 Instruction 0055-9017 6. In the example below, the current flow rate is 216.5 lpm, the background gas concentration i

Strona 27 - 2.17 Access Test IDs

Hi Flow Sampler Operation Instruction 0055-9017 33 5. The user is now asked to choose a new or different Test ID. Selecting Yes causes the Acces

Strona 28 - 2.17.3 Send Test IDs

Operation Hi Flow Sampler 34 Instruction 0055-9017 2.18.3 Leak Rate Measurement – Manual 2-Stage Mode Note: The Manual 2-Stage Mode can only be

Strona 29

Hi Flow Sampler Operation Instruction 0055-9017 35 7. Once a stable leak measurement is obtained, highlight Save and press the I/O ↵ key. Observe

Strona 30 - 2.17.4 Erase Test IDs

Operation Hi Flow Sampler 36 Instruction 0055-9017 2. Choose an attachment that will ensure the complete capture of the gas leak. Connect this att

Strona 31

Hi Flow Sampler Operation Instruction 0055-9017 37 8. The measurement process continues until Stop is selected. 9. Measurements that were sa

Strona 32

Operation Hi Flow Sampler 38 Instruction 0055-9017 Message Explanation Check Background Calibration Error – The background-gas sensor needs calib

Strona 33

Hi Flow Sampler Calibration Instruction 0055-9017 39 3 Calibration It is recommended that the Hi Flow Sampler be calibrated by your nearest Bacha

Strona 34

Table of Contents Hi Flow Sampler 4 Instruction 0055-9017 3.2 Gas Calibration Equipment Setup ...

Strona 35

Calibration Hi Flow Sampler 40 Instruction 0055-9017 Figure 3-1. Calibration Equipment Setup 3.3 Check Calibration Procedure This procedure chec

Strona 36

Hi Flow Sampler Calibration Instruction 0055-9017 41 7. Disconnect the hose from the GAS port, and then allow the pumps to run until both the Ba

Strona 37 - 2.19 Message Line

Calibration Hi Flow Sampler 42 Instruction 0055-9017 5. Apply the appropriate level of calibration gas from the calibration equipment to the instr

Strona 38 - ∇ ∇ ∇

Hi Flow Sampler Calibration Instruction 0055-9017 43 3.5 Calibration Amplifier Gain Check Use this procedure to display the gain values of the ca

Strona 39 - 3 Calibration

Maintenance Hi Flow Sampler 44 Instruction 0055-9017 4 Maintenance Routine maintenance of the Hi Flow Sampler consists of the following: • Char

Strona 40

Hi Flow Sampler Maintenance Instruction 0055-9017 45 Figure 4-1. Charging the Battery Pack 4.3 Removing the Cover When instructed to do interna

Strona 41 - Instruction 0055-9017 41

Maintenance Hi Flow Sampler 46 Instruction 0055-9017 4.4 Replacing the Internal Filters Three internal filters protect the instrument from being

Strona 42 - 42 Instruction 0055-9017

Hi Flow Sampler Maintenance Instruction 0055-9017 47 4.4.2 Leak-Gas Sensor Filter The leak-gas sensor filter is located inside the flow block of

Strona 43

Maintenance Hi Flow Sampler 48 Instruction 0055-9017 4.5 Gas Sensor Replacement Replace a gas sensor when it fails to calibrate per Section 3 Cal

Strona 44 - 4 Maintenance

Hi Flow Sampler Maintenance Instruction 0055-9017 49 4.6 Pump Replacement The pump located in either the leak-gas or background-gas sensor assemb

Strona 45 - 4.3 Removing the Cover

Introduction Hi Flow Sampler Instruction 0055-9017 5 1 Introduction 1.1 General Description and Use WARNING: The High Flow Sampler is not to be

Strona 46 - 4.4.1 Leak-Gas Filter

Maintenance Hi Flow Sampler 50 Instruction 0055-9017 4.7 Flow Rate Verification This procedure verifies the Hi Flow blower is operating at prope

Strona 47 - Instruction 0055-9017 47

Hi Flow Sampler Maintenance Instruction 0055-9017 51 Figure 4-8. Gas Flow System

Strona 48 - 4.5 Gas Sensor Replacement

Maintenance Hi Flow Sampler 52 Instruction 0055-9017 Figure 4-9. Plumbing Diagram

Strona 49 - 64" Allen Wrench

Hi Flow Sampler Maintenance Instruction 0055-9017 53 4.8 Setting Date and Time The following describes how to set the date and time as displayed

Strona 50 - 4.7 Flow Rate Verification

Maintenance Hi Flow Sampler 54 Instruction 0055-9017 3. Set the date by typing “DATWmm/dd/yy” where mm/dd/yy is the current month, day and year (e

Strona 51 - Figure 4-8. Gas Flow System

Hi Flow Sampler Troubleshooting Instruction 0055-9017 55 5 Troubleshooting 5.1 Instrument Repair It is recommended that field repair of the Hi F

Strona 52 - Figure 4-9. Plumbing Diagram

Troubleshooting Hi Flow Sampler 56 Instruction 0055-9017 6 Symptom 7 Probable Cause & Remedy Low flow rate reading with fully charged battery

Strona 53 - 4.8 Setting Date and Time

Hi Flow Sampler Troubleshooting Instruction 0055-9017 57 6 Symptom 7 Probable Cause & Remedy “Check Zeroing Bypass” message appears at bottom

Strona 54

Troubleshooting Hi Flow Sampler 58 Instruction 0055-9017

Strona 55 - 5 Troubleshooting

Hi Flow Sampler Parts & Service Instruction 0055-9017 59 8 Parts & Service 8.1 Replacement Parts Complete Hi Flow Sampler Assembly ...

Strona 56 - 56 Instruction 0055-9017

Hi Flow Sampler Introduction 6 Instruction 0055-9017 The main unit consists of an intrinsically safe, high-flow blower that pulls air from around

Strona 57

Parts & Service Hi Flow Sampler 60 Instruction 0055-9017 8.3 Service Centers United States Bacharach Sales/Service Center 621 Hunt Valley Cir

Strona 58 - Instruction 0055-9017

Hi Flow Sampler Declaration of Conformity Instruction 0055-9017 61 9 Declaration of Conformity DECLARATION OF CONFORMITY The manufactur

Strona 59 - 8 Parts & Service

Exclusive Distributor for the Hi Flow Sampler: 9030 Monroe Road, Houston, TX 77061 Ph: 713-844-1300

Strona 60 - 8.3 Service Centers

Introduction Hi Flow Sampler Instruction 0055-9017 7 1.4 Control Unit Description The control unit consists of an 8 line by 20 character LCD and

Strona 61 - 9 Declaration of Conformity

Hi Flow Sampler Introduction 8 Instruction 0055-9017 1.6 Technical Data TABLE 1-1. TECHNICAL SPECIFICATIONS Specification Description Informatio

Strona 62 - Printed in U.S.A

Hi Flow Sampler Operation Instruction 0055-9017 9 2 Operation 2.1 Operating Precautions To ensure that the gas sensors are properly zeroed at s