User’s Manual
Heath Consultants Incorporated
Houston, TX
Phone: 713-844-1300
Fax: 713-844-1309
1-800-HEATH-US
www.heathus.com
Heath....Safety, Leadership, Innovation, Performance
Then, Now and Tomorrow
Page 1 of 36 101515-0 RMLD MANUAL REV E
Przeglądanie stron 0
1 2 3 4 5 6 ... 37 38

Podsumowanie treści

Strona 1 - User’s Manual

™ User’s Manual Heath Consultants Incorporated Houston, TX Phone: 713-844-1300 Fax: 713-844-1309 1-800-

Strona 2

Transceiver The transceiver provides the laser launch for the infrared detection beam, visible green spotter laser, and spotter laser button. Harne

Strona 3 - Warnings

Optional Accessories Dual Shoulder Strap Harness The dual shoulder strap harness provides extra comfort and support for carrying the instrument for

Strona 4 - Chapter I

Chapter II Battery Charging Rechargeable Battery The RMLD has an internal, rechargeable Li-ion battery that provides the main power to the instrum

Strona 5

Battery Charger The RMLD is provided with a universal AC battery charger. The plug of the charger can be changed to fit the type of receptacle used

Strona 6 - Overview

Chapter III Operating the RMLD This chapter provides information on the use of the RMLD. It provides an explanation of working with the menu, set

Strona 7

Headphone port Charger port RS-232 port External Power port Fig. 3-2: RMLD rear panel Headphone port: receptacle for the head

Strona 8

Turning On the Spotter Laser **Caution** The green spotter laser in under the control of the operator and is on only when the spotter trigger switc

Strona 9 - RMLD System Components

By pressing the Select button, the Menu options are scrolled through in the following order: • Self-Test i. Press Up button to initiate the Self-T

Strona 10 - 101515-0 RMLD MANUAL REV E

• BACLI i. Press Up button to turn on or to turn off the display backlight. Fig. 3-5 Although the ppm-m readout will not be updating, the

Strona 11 - Optional Accessories

warning is sounding, the RMLD may still be able to detect very large gas concentrations, indicated by a high pitch tone. The Alarm Detection Thresho

Strona 12 - Chapter II

Notice The contents of this user’s manual are proprietary to Heath Consultants Incorporated (HEATH). Reproduction in whole or in part of this manua

Strona 13

5. If the location with the highest pitch is consistent then the location of the leak is at that spot. 6. If the pitches are not consistent then

Strona 14 - Chapter III

Press the UP button to return to the beginning of the SELF-TEST and refer to the next section: Calibration Override. 7. Press the UP button. This w

Strona 15 - **Note**

When the Self-Test fails. The following table may help to isolate the problem: Cause Corrective Action Low signal strength or noisy • Make sure ba

Strona 16

Chapter IV Surveying with the RMLD A training video is provided with the RMLD, which demonstrates the best overview of techniques used. It is impo

Strona 17

• Move in closer if the range is too far or ground elevation causes the beam to not come into contact with the ground (dark zones). While scanning

Strona 18 - Using the DMD Mode

Be aware of the ground elevation. Scanning across the top of a knoll or past the edge of a structure can result in beam skips (a sudden change in di

Strona 19 - Using the Pure Tone Mode

Dealing with False Detections While using the DMD mode, several conditions may occur that will cause the algorithm to give a detection indication.

Strona 20 - Self-Test and Calibration

D i r e c t S u n a n d A m b i e n t L i g h t A d d s D e t e c t e d B a c k g r o u n d P h o t o n s

Strona 21 - Calibration Override

Fig. 4-5: The ppm-m reading is the product of the plume concentration times the distance the IR beam passes through the plume (plume width) in addi

Strona 22 - **Caution**

Chapter V Maintenance and Troubleshooting Information Troubleshooting the Instrument The advanced design of the RMLD makes it one of the most rel

Strona 23 - Chapter IV

Warnings It is essential that users of this instrument read, understand, and follow the instructions for operation and maintenance, and the precauti

Strona 24 - Long Range Scanning:

• Low battery recharge if necessary Excessive false detections while scanning at longer distances • Scanning too fast • Alarm detection thresho

Strona 25

Maintenance In order to maintain the RMLD in good working condition, the following maintenance should be performed as indicated: Maintenance Item F

Strona 26 - Detector

Chapter VI Service Information Warranties and Warranty Repair The RMLD is warranted to be free from defects in material and workmanship for one (1

Strona 27

Glossary Beam Skip: Occurs when the IR beam jumps between a near object and a far away object. This may cause a false detection. Control Module:

Strona 28

Page 34 of 36 101515-0 RMLD MANUAL REV E

Strona 29 - Chapter V

Page 35 of 36 101515-0 RMLD MANUAL REV E

Strona 30

RMLD Daily Self-Test and Calibration Log RMLD S/N Date Operator Passed Self-Test? Self-Test ppm-m Reading Alarm Level Notes

Strona 31 - Maintenance

Declaration of Conformity Manufacture’s Name: Heath Consultants Manufacture’s Address: 9030 Monroe Road Houston, TX 77061 Model: Remote

Strona 32 - Chapter VI

Technical Support 1-800-HEATH-US (1-800-432-8487) HOUSTON FACTORY SERVICE CENTER Heath Consultants Manufacturing Division 9030 Monroe Road H

Strona 33 - Glossary

Table Of Contents Notice...……….. 2 Warnings...

Strona 34

Chapter IV Surveying with the RMLD……………………………… 23 Long Range Scanning…………………………………… 24 Dealing with False Detections…………………………… 26 How Does the RM

Strona 35

Overview The HEATH Remote Methane Leak Detector (RMLD™) is a highly advanced technology, capable of detecting methane leaks from a remote distance.

Strona 36

The RMLD consists of two main interactive components: • Transceiver • Control Module Fig 1-3: The RMLD system consists of a Laser Emitter/Receiv

Strona 37 - Declaration of Conformity

Chapter I RMLD System Specifications Detection Method: Tunable Diode Laser Absorption Spectroscopy (TDLAS) Measurement Range: 0 to 99,999 ppm-m Sens

Strona 38

RMLD System Components This Chapter describes the features of the RMLD. Please refer to Figure 1-4 for illustrations of individual parts.