TEST / CALIBRATION EXAMPLE
Fig. 6-9 Setup Continuation
11.      
corresponding value should be entered in ‘%’ window.
The value on the gas cylinder label in this example is 50%.
To enter this (50) value, highlight existing (0) value then type
new value in window.
12.  to display
gases drop down list, as illustrated in Fig. 6-6.


To enter this (100) value, highlight existing (0) value then
type new value in window.
13. To s p eci f y g as No. 5 ( 0 to 10 0 0pp m C O ), s el e ct to
display gases drop down list, as illustrated in Fig. 6-6.

value on the gas cylinder label in this example is 500 PPM.
To enter this (500) value, highlight existing (0) value then
type new value in window.