Users Manual
Natural Gas (Methane)
ODORATOR
®
Przeglądanie stron 0
1 2 3 4 5 6 ... 30 31

Podsumowanie treści

Strona 1 - ODORATOR

User’s ManualNatural Gas (Methane)ODORATOR®

Strona 2 - PROPRIETARY NOTICE

Operation1OPERATION:This Chapter discusses the proper steps to safely operate the ODORATOR. Also included are procedures for periodic testing and iden

Strona 3

Operation2. Connect the sample hose to the gas supply outlet valve and to the ODORATOR inlet port (1).3. Open the gas supply outlet valve and the OD

Strona 4

Operation12. If the exhaust does not have a readily detectable odor repeat step 11. If the exhaust does have a readily detectable odor, push and hol

Strona 5 - INTRODUCTION

Operation41. Gas Inlet 6. Calibration Sticker 2. Power Switch 7. Flow Metering Valve3. Power Indic

Strona 6

Operation5Periodic Test:1. Every thirty (30) days the ODORATOR should be operated with the power switch on and while not hooked up to a gas supply.2.

Strona 7 - WARNINGS AND CAUTION

Maintenance6Troubleshooting ChartMAINTENANCE:“LO BATT” appears on LCD.Weak batteries. Replace batteries.Motor and LCD will not turn on.RMASpeed Indica

Strona 8

Maintenance7Parts ListThe following list species ODORATOR parts and accessories that are available from the factory or regional ofce. 07156

Strona 9 - TABLE OF CONTENTS

Appendix A8ActionResults Operator NameTest Date Periodic Test Record

Strona 10 - Chapter I

Appendix AActionResults Operator NameTest Date Periodic Test Record9

Strona 11

Appendix B10Appendix BODORATOR Correction Chart for MethaneThis chart assumes the ODORATOR has been properly zeroed and allowed to run for at least on

Strona 12

PROPRIETARY NOTICEThe contents of this instruction manual are proprietary to the Manufacturing Division of Heath Consultants Incorporated. Reproducti

Strona 13 - Side View

Appendix C11Appendix CODORATOR Use at ElevationThe ODORATOR works by allowing a metered amount of sample gas to mix with a xed amount of air and the

Strona 14 - Chapter II

Appendix C12**Note** The elevation correction chart supports new style Meth-ane ODORATORS only. Instruments manufactured before July 1997, that have

Strona 15 - Chapter III

Appendix C13Elevation Corrections for the Methane ODORATORThe methane ODORATOR mixes air with natural gas to allow the user to determine the percent g

Strona 16

Appendix CThe reading as a percentage between the adjacent chart readings can be used to determine the actual value using the same percentag

Strona 17 - Periodic Test Record

Appendix CTherefore the actual value will be 30% of the difference in the chart higher actual from the chart lower actual. Again, the lower value will

Strona 18

Appendix C16Elevation Change Correction ChartInstructions for correction on page 13Down 5,000’Down 4,000’Down 3,000’Down2,000’Down 1,000’Up 1,000’

Strona 19 - Appendix B

Service17SERVICE INFORMATION:Warranties and Warranty RepairAll instruments and products manufactured by Heath Consultants Incorporated are warranted

Strona 20 - ODORATOR Use at Elevation

Service18SERVICE INFORMATION:Return Procedure (RMA):The following steps will expedite the repair of your instrument:1. Contact Heath Factory Service

Strona 21

Service19SALES, CUSTOMER SERVICE, AND INSTRUMENT REPAIR National Toll Free # 1-800-HEATH-US (1-800-432-8487)REGIONAL SALES, CUSTOMER SERVICE, AND INST

Strona 22

Service20NOTES

Strona 23

ODORATOR®User’s ManualHeath Consultants IncorporatedHouston, TX713/844-1300 Fax: 713/844-13091-800-HEATH-USwww.heathus.comHeath...Safety, Leadershi

Strona 24

Service21NOTES

Strona 25

Heath...Safety, Leadership, Innovation, PerformanceThen, Now and TomorrowHeath Consultants IncorporatedHouston, TX713-844-1300Fax: 713-844-13091-800-

Strona 27

iNatural gas (methane) to be tested enters the ODORATOR through the barbed tting at the gas supply inlet port. Internally, the gas passes through an

Strona 28

iiWhen the ODORATOR is turned on, the user might momentarily notice the letter “L” on the LCD when the READ switch is depressed and held down. I

Strona 29

WARNINGS AND CAUTIONPRIOR TO USE, OPERATOR SHOULD READ ALL PRECAUTIONS, INSTRUCTIONS FOR OPERATION, AND MAINTENANCE REQUIREMENTS CONTAINED IN PRODUCT

Strona 30

ivANY TIME GAS IS BEING PASSED THROUGH THE ODORATOR THE POWER SWITCH MUST BE TURNED ON. THIS WILL DILUTE THE GAS AT THE EXHAUST PORT AND PREVENT POCKE

Strona 31 - 1-800-HEATH-US

INTRODUCTION ...................iWARNINGS AND CAUTION .........