Excessive false detections
while scanning at longer
distances
• Alarm detection
threshold set too low
• Scanning at the range
limit of the instrument
scanning rate. Pause at
the long range and
sweep towards you
• Increase the Alarm
detection threshold
•
Excessive false detections
while scanning at closer
distances
• Alarm detection
threshold set too low
motions while scanning
• Increase the Alarm
Excessive false detections
or loss of sensitivity
optimized
• Alarm Detection
Threshold set too high
procedure
• Check the Alarm
Detection Threshold
Error icon or Warning icon
on continuously
• Moisture condensation
on mirror due to rapid
change in temperature
• Internal component
failure
• Check battery level and
recharge if necessary
• Allow for the
temperature of the
instrument to stabilize
• Note error code and
Battery indicator does not
show full charge after
charging
charged
• Battery level
calibration error
• Charge unit until solid
green light on charger
is on
• Run unit until it shuts
off, then fully recharge
with out interruption
Battery indicator bars are
flashing on and off
calibration error
• Run unit until it shuts
off, then fully recharge
101296-0 RMLD IS MANUAL REV J Page 32 of 40